Magnetic Component in E-Cigarettes Found to Interfere with Implantable Cardioverter-Defibrillator Function

Potential for unintended temporary programming and arrhythmic complications when an electronic cigarette is placed in close proximity to an implantable cardioverter-defibrillator or pacemaker reported in HeartRhythm Case Reports.

2020 Press Release/Statements

Philadelphia, March 16, 2020 – An e-cigarette carried in the left breast shirt pocket of a patient with an implantable cardioverter-defibrillator (ICD) caused magnetic reversion, interrupting the ICD's ability to detect and treat dangerous heart rhythm problems, clinicians report in Heart Rhythm Case Reports, published by Elsevier. The patient was not aware that the e-cigarette has an integrated magnetic component, and it had suspended detection of heart rhythm problems by the ICD four times before he reported it to his healthcare team.

"To our knowledge this is the first reported case of magnetic reversion of an ICD by an e-cigarette," stated senior investigator Usha B. Tedrow, MD, MPH, and lead authors Julie B. Shea, MS, RNCS, Martin Aguilar, MD, and William Sauer, MD, from the Cardiovascular Arrhythmia Service at Brigham and Women's Hospital and Harvard Medical School, Boston, MA, USA. "Given the increasing use of e-cigarettes worldwide, recognition of this potentially serious interaction appears clinically important."

A 48-year-old male with an ICD reported that he heard his device "beep" several times, both at home and at his office. There were no adverse symptoms associated with the beeping, and remote monitoring found the device was working normally. The patient denied any magnetic exposure, but data provided by the ICD manufacturer found four instances of magnet interactions with the device, corresponding with when the patient heard the tone.

Upon further questioning, the patient recalled using his e-cigarette (JUUL vape device), which he frequently stored in his left breast pocket, overlying the ICD. When the healthcare team held the device up to his ICD, it triggered the steady magnet tone. The patient was educated about the importance of keeping any type of magnet away from his ICD. JUUL's website does recommend keeping e-cigarettes away from key cards, credit cards, and other items with magnetic strips, as well as pacemakers.

"Magnets are ubiquitous in commercially-available electronic devices. They can be integrated in ways that are difficult to recognize. Although manufacturers are not routinely required to specify the strength of the magnetic fields and safety information for interference with medical-grade devices, the general recommendation is that any portable electronic or magnetic device be kept at least six to 12 inches away from an implant," noted Dr. Tedrow and Ms. Shea.

There are commercially available magnetic field meters, and even several smartphone applications, that can be used to estimate the strength of a magnet. Practically speaking, most cardiac implantable devices have a magnetic exposure upper limit of 10G, and manufacturers typically recommend a 2:1 safety margin for safe clinical operation. "As such, finding the distance at which the magnetic field is 5G or less would, in principle, provide adequate clearance for safe clinical operation of cardiac implantable devices," Dr. Aguilar added.

"Practitioners should remain vigilant regarding the use of new technology by their ICD patients. In our case, there was no adverse effect from the interaction of the e-cigarette with the device, but if it had happened during a tachycardia episode, it could have had serious, perhaps even fatal consequences," cautioned Dr. Tedrow and Ms. Shea.

###

Notes for editors
The article is "Unintentional magnet reversion of an implanted cardiac defibrillator by an electronic cigarette," by Julie B. Shea, MS, RNCS, FHRS, Martin Aguilar, MD, William H. Sauer, MD, FHRS (EP Division Chief), and Usha Tedrow, MD, MPH, FHRS (https://doi.org/10.1016/j.hrcr.2020.01.013). It appears in HeartRhythm Case Reports, volume 6, issue 3 (March 2020) published by Elsevier. It is openly available at www.heartrhythmcasereports.com/article/S2214-0271(20)30017-8/fulltext.

Full text of this article is also available to credentialed journalists upon request; contact Jane Grochowski at +1 406 542 8397 or [email protected]. Journalists who wish to interview the authors should contact Mark Murphy, Office of Strategic Communication, Brigham and Women's Hospital, at +1 617 732 9551 or [email protected].

###

About Heart Rhythm Case Reports

Heart Rhythm Case Reports provides rapid online electronic publication of the most important current case reports, illustrations, and educational vignettes in the field of cardiac arrhythmias and electrophysiology. The Journal publishes case reports and series devoted to the diagnosis and treatment of heart rhythm disorders, as well as the electrophysiology of the heart and blood vessels. All articles are peer-reviewed. The Journal is published online only with open access. www.heartrhythmcasereports.com

About the Heart Rhythm Society

The Heart Rhythm Society is the international leader in science, education, and advocacy for cardiac arrhythmia professionals and patients, and the primary information resource on heart rhythm disorders. Its mission is to improve the care of patients by promoting research, education, and optimal healthcare policies and standards. The Heart Rhythm Society is the preeminent professional group representing more than 5,100 specialists in cardiac pacing and electrophysiology from more than 70 countries. www.HRSonline.org 

About Elsevier

Elsevier is a global information analytics business that helps scientists and clinicians to find new answers, reshape human knowledge, and tackle the most urgent human crises. For 140 years, we have partnered with the research world to curate and verify scientific knowledge. Today, we’re committed to bringing that rigor to a new generation of platforms. Elsevier provides digital solutions and tools in the areas of strategic research management, R&D performance, clinical decision support, and professional education; including ScienceDirect, Scopus, SciVal, ClinicalKey, and Sherpath. Elsevier publishes over 2,500 digitized journals, including The Lancet and Cell, 39,000 e-book titles and many iconic reference works, including Gray’s Anatomy. Elsevier is part of RELX, a global provider of information-based analytics and decision tools for professional and business customers. www.elsevier.com

Membership Reminder

RENEW MEMBERSHIP